When it comes to solvents, what's the difference between HPLC and LC-MS grade? Does it matter? If so, how?
In short, yes. There is a distinct difference between solvent grades when it comes to HPLC and LC-MS purposes. Even though both systems require the extremely low concentrations of impurities for their analyses, LC-MS (Liquid Chromatography-Mass Spectrometry) requires an even higher level of purity than HPLC (High-Performance Liquid Chromatography) due to higher sensitivity to a wider range of contaminants. Because of this, LC-MS-grade solvents may work just fine for HPLC purposes, although the purity is higher than the necessary level needed for liquid chromatography. On the other hand, HPLC-grade solvents may affect and reproducibility of test results and damage equipment if used in LC-MS or a mass spectrometer.
Since mass spectrometers and their detectors are highly sensitive to extremely small traces of impurities or contaminants, LC-MS solvents have more stringent purity requirements than HPLC and other analysis techniques. While the absolute difference in purity may seem small, as both require well over 99.9% purity, the relative difference in purity requirements can span orders of magnitude. Depending on the impurity, For HPLC experiments, contaminants at concentrations in the low parts per million range are acceptable, but the impurity threshold for LC-MS solvents is often below 10 parts per billion, depending on the impurity. For context, 10 parts per million of impurities corresponds to 99.999% pure, while 10 parts per billion of impurities is 99.999999% pure!
Fundamental differences between HPLC and LC-MS
Liquid Chromatography-Mass Spectrometry (LC-MS) combines the separation power of HPLC with the detection capabilities of mass spectrometry. LC-MS allows for sensitive detection and molecular identification, which is useful for more complex sample analysis. Because LC-MS integrates mass spectrometry for additional analysis, it provides structural information about molecules by analyzing the mass of ionized compounds and fragments. LC-MS technology enables detailed molecular weight measurements, making it suitable for complex biological samples.
Liquid Chromatography-Mass Spectrometry (LC-MS) combines the separation power of HPLC with the detection capabilities of mass spectrometry. LC-MS allows for sensitive detection and molecular identification, which is useful for more complex sample analysis. Because LC-MS integrates mass spectrometry for additional analysis, it provides structural information about molecules by analyzing sample mass. LC-MS technology enables detailed molecular weight measurements, making it suitable for complex biological samples.
Differentiating Solvent Grades
High-Performance Liquid Chromatography (HPLC) and Liquid Chromatography-Mass Spectrometry (LC-MS) solvents are specifically formulated to meet the rigorous needs of their respective techniques. HPLC solvents must have minimal non-volatile residue. They are filtered through 0.2 or 0.45-micron or smaller filters to eliminate harmful particles. These solvents must be at least 99.9% pure to avoid interfering with the analysis. Detectors in HPLC are less sensitive to impurities that do not have strong UV absorbance.
In contrast, LC-MS solvents need even higher purity levels, often surpassing HPLC standards by removing additional impurities that could affect the mass spectrometry detector. They typically undergo further testing, such as the reliable reserpine test, which is run in both electrospray ionization (ESI) and atmospheric chemical pressure ionization (APCI) tests to ensure high purity with only a few billionths of a percentile in impurities.
Solvent Purity Impacts on LC-MS Interferences
While it’s clear that a general laboratory-grade solvent isn’t appropriate for highly sensitive analytical techniques, there are still questions around whether high-performance liquid chromatography (HPLC) grade solvents can be used instead of LC-MS grade. In reality, this choice can have significant impacts, as LC-MS-grade solvents go through extra purification steps to remove trace metals and other impurities that could interfere with analysis.
Particulate matter can block inlet and in-line filters, causing technical issues with instruments such as column life span reduction and machinery downtime. This makes it crucial to select LC-MS solvents pre-filtered with a 0.2 μm (or smaller) filter to eliminate these impurities.
Solvents must be free of impurities with UV-vis absorbance, such as organic acids. LC-MS systems often incorporate diode-array detection, making a stable, flat baseline with minimal drift essential. Likewise, solvents with minimal mass background are critical for trace MS applications, enhancing signal intensity at low analyte concentrations. Contaminants like plasticizers or phthalates can enter during manufacturing—from surfaces or equipment in production areas—so manufacturers must rigorously control and monitor these potential sources of residuals.
Metal adduct ion formation presents a specific challenge in ESI analyses, where alkali metals like sodium and potassium associate with ion fragments of the analyte. Adduct formation is less likely with volatile ammonium salts. Adduct formation has two primary effects: first, it complicates MS spectra by changing the mass-to-charge ratio of the analyte ions and fragments. Second, adduct formation can vary between runs, potentially compromising the accuracy and reproducibility of analytical results. To minimize these issues, selecting a solvent that has been rigorously tested for metal ions and other ionic contaminants that may cause ion suppression is essential.
Benefits of HPLC grade solvents
While LC-MS applications require solvents that met HPLC purity standards, certain factors are crucial specifically for mass spectrometry applications that may be less important for HPLC-only applications:
-
Ultra-low Metal Ion Content: LC-MS solvents are purified to reduce metal ions like sodium, potassium, or iron that can interfere with MS sensitivity and cause unwanted adduct formation. For HPLC, such rigorous reduction of metal ions isn't usually required, as metal adducts don’t affect UV or fluorescence detection. However, in specific HPLC applications involving chelating agents or ion-pairing methods, metal ions may still impact performance.
-
Minimized Mass Background: LC-MS solvents undergo extra filtration to reduce any background ions that may appear on the MS spectrum. This level of background control is unnecessary for HPLC, where detection is influenced by UV-absorbance.
-
Volatile Contaminants Control: LC-MS solvents are specifically refined to minimize volatile organic contaminants that could interfere with MS readings. These contaminants are less of an issue for HPLC, where volatility doesn’t directly influence detection methods like UV or PDA, unless the volatile contaminant has strong UV-absorption.
-
Ion Suppression and Enhanced Purification for Low Detection Limits: LC-MS analyses are often conducted at very low analyte concentrations, requiring solvents that minimize ion suppression to maintain signal intensity. In HPLC, ion suppression is not a concern because the detection methods do not rely on ionization, so stringent ion suppression measures aren’t necessary.
-
Consistency in Baseline Stability for MS: LC-MS grade solvents are designed to provide a stable baseline in mass spectrometry, minimizing noise due to impurities. HPLC doesn’t require this level of consistency, as its baseline stability is generally unaffected by the same level of trace contaminants.
Because LC solvents without MS applications do not require these specifications, other factors such as ease of use and cost-effectiveness make HPLC solvents more attractive for their specific purpose. Since liquid chromatography excels in separation capabilities, HPLC-grade solvents are an optimal and economic choice for their application.
Common Solvents Used in HPLC and LC-MS
In both HPLC and LC-MS methods, several solvents are commonly used. Acetonitrile and methanol are frequently chosen due to their compatibility and effectiveness. In LC-MS, protic solvents such as mixtures of methanol and water offer good ionization efficiency and are widely used in solvent selection.
When paired with additives, LC-MS grade water can enhances separation and detection. Another consideration is solvent gradients; designing these carefully can improve the separation, chromatographic resolution, and efficiency of your analysis. Understanding and selecting the appropriate solvent for your specific analysis can significantly impact the outcome.
Impurities Found in Different Solvents
When choosing solvents for HPLC and LC-MS applications, it's important to consider impurities that can affect the results. Different solvents may contain various impurities that can alter their performance. Understanding these impurities helps you choose the best solvent for your needs.
-
Water: Water used in chromatography should be free from ions and organic compounds. Impurities may include ions like calcium and magnesium, which can interfere with analyses.
-
Methanol: This solvent can contain trace metal impurities such as aluminum, barium, and cadmium. These metals can affect the sensitivity of your analysis and lead to unwanted peaks and enhance background noise.
-
Acetonitrile: In acetonitrile, you may encounter metal ions, with aluminum and calcium being the most common. These impurities can lead to fluctuations in the baseline and inaccurate results.
-
Isopropyl Alcohol: Common impurities in isopropyl alcohol include peroxides and aldehydes. These can cause degradation of samples or changes in the retention times of analytes.
-
Dichloromethane: This solvent may contain stability-related impurities like hydrochloric acid and water. These contaminants can cause issues with column longevity issues and increased baseline noise.
-
Tetrahydrofuran (THF): THF often contains peroxides, especially if not stored properly. Peroxide accumulation can lead to hazards in the lab and impact the degradation of analytes during separation.
Selecting high-purity solvents with minimal impurities ensures better accuracy and reliability in your chromatography experiments.
Conclusion:
Given these factors, HPLC can be very effective for routine analysis and situations where you do not need extensive molecular identification. It stands as a valuable method in various laboratories for these practical reasons. On the other hand, LC-MS combines the strengths of HPLC with additional tools for structural elucidation and detection. This combination results in a versatile and powerful analytical technique, suitable for a wide range of research and industrial applications.
Quick Comparison:
Feature |
HPLC |
LC-MS |
---|---|---|
Purpose |
Separation of compounds |
Separation and molecular identification |
Sensitivity |
Moderate |
High, useful for trace-level detection |
Applications |
Chemical analysis |
Pharmaceuticals, environmental pollutants, and biological laboratories |
Cost |
Lower Equipment Costs |
Higher Costs |
Ease of Use |
Less Complex Operation |
More Complex |
Sample Integrity |
No Ionization Required |
Ionization Required |
Sources for this Article
-
https://www.birchbiotech.com/collections/hplc-grade/products/dichloromethane-hplc-grade
-
https://www.birchbiotech.com/collections/lc-ms-grade/products/isopropyl-alcohol-lc-ms-grade
-
https://www.birchbiotech.com/collections/lc-ms-grade/products/methanol-lc-ms-grade-1
-
https://www.birchbiotech.com/collections/hplc-grade/products/tetrahydrofuran-hplc-grade
-
https://www.birchbiotech.com/collections/lc-ms-grade/products/water-lc-ms-grade
Disclaimer: The content provided on the Birch Biotech blog is for educational and entertainment purposes only. The information offered here is designed to provide helpful insights and advice related to laboratory practices and supplies.
Readers are advised to refer to our product-specific quality data sheets and Certificates of Analysis (COAs) available on our website for detailed information on product specifications. It is essential to handle and store all materials according to the safety guidelines and regulatory requirements applicable to your area.
While we endeavor to ensure the accuracy and relevance of the information published, it should not be used as a substitute for professional advice or official protocols. We encourage all our readers to consult their institution's guidelines, local regulations, and professional standards before implementing any practices discussed here.
Birch Biotech does not accept liability for any actions undertaken based on the information provided in this blog nor for the misuse of our products. Furthermore, Birch Biotech does not guarantee the completeness, reliability, or timeliness of the information contained on this website.
This disclaimer is subject to change at any time without notifications.